Table 2.
2a. Test Characteristics of the 3D-CAM | |||
Delirium Diagnosis (− or +) | Reference Standard (+) | Reference Standard (−) | 3D-CAM Totals |
3D-CAM (+) | 18 | 6* | 24 |
3D-CAM (−) | 1 | 76 | 77 |
Reference standard totals | 19 | 82 | 101 |
Test characteristic (%) | Sensitivity = 18/19 (95 %) | Specificity = 76/82 (93 %) | |
95 % CI | [74 %,100 %] | [85 %, 97 %] | |
2b. Test Characteristics of the CAM-ICU | |||
Delirium Diagnosis (− or +) | Reference Standard (+) | Reference Standard (−) | CAM-ICU Totals |
CAM-ICU (+) | 10 | 0 | 10 |
CAM-ICU (−) | 9 | 82 | 91 |
Reference standard totals | 19 | 82 | 101 |
Test characteristic (%) | Sensitivity = 10/19 (53 %) | Specificity = 82/82 (100 %) | |
95 % CI | [29 %,76 %] | [96 %,100 %] |
CI confidence interval
*Of the six 3D-CAM false positives, four were judged to have subsyndromal delirium based on the reference standard assessment