Skip to main content
. 2014 Apr 30;8(4):691–698. doi: 10.1177/1932296814533172

Table 4.

Summary of Patient Blood Volume Collection.

All lancing eventsa One lancing eventb Repeat lancing eventc
No. of patients 64 53 11
No. of fingersticks 512 424 88
Blood volume, µL
 Mean ± standard deviation 3.1 ± 2.1 3.1 ± 2.0 2.9 ± 2.4
 Median 2.7 2.7 2.3

Each patient performed a minimum of 8 fingersticks (lancing events). Eleven patients required second lancing attempts to express ≥0.3 µL. Volume data were recorded only for the second fingerstick attempt in these patients.

a

The mean blood volume per patient was used to calculate the mean and median blood volume across all 64 patients.

b

Mean and median blood volume calculated using all 424 fingersticks.

c

Mean and median blood volume calculated using all 88 fingersticks.