. 2014 May 12;8(4):766–775. doi: 10.1177/1932296814532575

Table 2.

Accuracy Results of 100-Participant Subset for Each of 4 Systems Tested According to ISO15197: 201316 Criteria.

DIN ISO15197:2013
MARD
Overall within accuracy limits
<100 mg/dl
>100 mg/dl
±5 mg/dl ±10 mg/dl ±15 mg/dl ±5% ±10% ±15%
System % n % % % % % % %
Control 11.6 156/200 78.0 29.5 68.2 84.1 2.6 33.3 76.3
Noncorrected prototype 9.6 186/200 92.5 38.6 72.7 97.7 32.1 62.2 80.1
Corrected prototype 4.9 199/200 99.5 61.4 81.8 97.7 59.6 94.9 100