Skip to main content
. 2014 Apr 21;8(4):709–719. doi: 10.1177/1932296814531356

Table 4.

Paired CGM sensor readings in the Clarke error grid zones in hypoglycemia.

Hypoglycemia (%)
Euglycemia (%)
Hyperglycemia (%)
Full glycemic range (%)
Calibration algorithm Zone A Zone D Zone E Zone A Zone B Zone C Zone A Zone B Zone C Zone D Zone E Zone A+B
2-point calibration 77.5 21.3 1.2 73.1 26.1 0.8 79.9 16.4 0.1 2.5 1.1 96.6
1-point calibration 91.6 7.3 1.1 73.4 25.8 0.8 83.2 13.2 0.1 2.1 1.4 97.7