. 2016 Feb 25;54(3):620–624. doi: 10.1128/JCM.02778-15

TABLE 1.

Performance of WASPLab digital imaging compared to manual reading

Clinical test site No. of specimens tested Results (no.)a
Performance (% [95% CI])b
MP/AP MN/AN MN/AP MP/AN Sensitivity Specificity
1 5,604 119 5,266 219 0 100 (96–100) 96.0 (95–96)
2 41,064 680 36,333 4,051 0 100 (99–100) 90.0 (89–90)
3 2,217 162 1,898 157 0 100 (97–100) 92.4 (91–93)
4 8,805 406 7,616 783 0 100 (99–100) 90.7 (90–91)
Total 57,690 1,367 51,113 5,210 0 100 (99–100) 90.7 (90–91)
a

MP/AP, manual positive/automation positive; MN/AN, manual negative/automation negative; MN/AP, manual negative/automation positive; MP/AN, manual positive/automation negative.

b

CI, confidence interval.