. 2016 Feb 25;54(3):625–630. doi: 10.1128/JCM.02876-15

TABLE 2.

PCR results of the 147 samplesa and concordance rate between the four PCR methods

PCR method and result Amplisens PCR
MycAssay PCR
Bio-Evolution PCR
Positive Negative Positive Negative Positive Negative
In-house PCR (11)
    No. positive 77 1 73 5 55 23
    No. negative 5 64 0 69 0 69
    Kappa concordance coefficient (95% CI) 0.92 (0.85–0.98) 0.93 (0.87–0.99) 0.69 (0.58–0.80)
    Concordance, % 95.9 96.6 84.4
Bio-Evolution PCR
    No. positive 55 0 55 0
    No. negative 27 65 18 74
    Kappa concordance coefficient (95% CI) 0.64 (0.53–0.76) 0.75 (0.65–0.86)
    Concordance, % 81.6 87.8
MycAssay PCR
    No. positive 72 1
    No. negative 10 64
    Kappa concordance coefficient (95% CI) 0.85 (0.77–0.93)
    Concordance, % 92.5
a

There is one missing result for the MycAssay PCR.