Skip to main content
. 2016 Feb 24;43(3):1411–1420. doi: 10.1118/1.4941953

TABLE IV.

Summary of CT number performance for each reconstruction kernel provided with FreeCT_wFBP. All values are within acceptable ACR ranges. All values are in HU.

Polyethylene Bone Acrylic Air Water
Smooth −88.3 869.3 123.6 −989.1 −0.4
Medium −88.3 869.0 123.5 −988.7 −0.4
Sharp −88.1 866.6 123.4 −988.1 −0.3
Acceptable range −107 to −84 850 to 970 110 to 135 −1005 to −970 −7 to 7