Skip to main content
. Author manuscript; available in PMC: 2016 Feb 29.
Published in final edited form as: Magn Reson Med. 2015 Feb 17;75(1):441–451. doi: 10.1002/mrm.25587

Table 1.

Synergies Between RF Receive Arrays and Multi-Coil Shim Arrays

Design Principle Benefit for RF Receive Array Benefit for MC Shim Array
Place loops as close to body
 as possible.
Increase B1 sensitivity and SNR. Generate ΔB0 shim field in body with
 high efficiency.
Use as many coils/degrees
 of freedom as practical.
Improve parallel imaging performance. Cancel higher-order B0 inhomogeneity.

MC, multicoil; RF, radio frequency; SNR, signal-to-noise ratio.