Skip to main content
. Author manuscript; available in PMC: 2016 Mar 1.
Published in final edited form as: Nat Protoc. 2006;1(6):3129–3142. doi: 10.1038/nprot.2006.358

Table 1.

Troubleshooting table.

Problem Possible reason Solution
No signal Quality of metaphases
Slides are over 1 year old
If there is a cell pellet stored, try dropping new slides with better humidity conditions and/or longer hypotonic conditions
Excess cytoplasm surrounding the chromosomes Try a stronger pepsin treatment; generally, the pepsin stock solution (slide pretreatment solution 2) can be increased in 5-μl increments
Probe denaturation temperature Check the actual temperature of the equipment used to denature the probe
Slide denaturation temperature Check the actual temperature of the instrument used to heat the slides (water bath or hot plate). Always check the structural integrity of the slide after pretreatment and denaturation (before hybridization); chromosomes should not appear hollow or too shiny
Detection reagents are approaching expiration Always check the date on the detection reagents, as some expire within 3 months of purchase
Detection reagents were improperly diluted It is important to not overdilute the detection reagents; check the concentrations
Antibodies were added in the wrong order It is important to first add the mouse anti-digoxygenin and then detect that with the anti-mouse fluor
The wrong filter was used during imaging Always make sure to use the SKY filter cube for taking SKY images.
Weak signal The chromosomes were too phase-light from the start When checking the slide, make sure to choose an area of hybridization that contains darkly stained chromosomes, not light gray ones. If no dark areas exist, try to drop new metaphases at a higher humidity
Forgot to use RNase and/or pepsin These steps remove excess cytoplasm and RNA, allowing the probe to make full contact with the DNA on the slide (see Box 1)
Preannealing time was too extensive Longer preannealing times allow for unique single-copy genomic sequences to form double-stranded DNA with their complementary strands in the tube, rather than on the slide.
Wash solutions were stored a long period of time Always make up fresh solutions, especially wash 1. Make sure the pH is in the range of 7–7.4
High background/low signal specificity Detection was not performed at the correct temperature The stringency of the washes can be compromised by a lower temperature; therefore, check the actual temperatures of the water baths
Slides were not denatured long enough Always check the appearance of chromosomes after denaturation; they should look as good as when you started the procedure. Try increasing time in 5- to 10-s increments
Insufficient hybridization time Try a 72-h incubation
Excess cytoplasm surrounding the chromosomes See Box 1
BSA was not properly washed off Always check the BSA concentration when making wash solution 4. Also note the temperature of the wash solutions after Step 42