TABLE III.

Summary of particle size measured by the DiSCmini (DM) and the SMPS reference instrument for tests with monodispersed aerosols.

Test Aerosol Composition dDMA Concentration Range* DM SMPS Size Ratio (rs) P-value**
Avg. Geo. Mean Dia. (Std. Dev.) nm Avg. Geo. Mean Dia. (Std. Dev.) nm Avg. Geo. Std. Dev.
NaCl 30 L 29 (3.8) 31 (0.44) 1.4 0.94 0.33
M 46 (1.6) 36 (0.51) 1.6 1.28 <0.01
H - - - - -
100 L 105 (1.7) 107 (0.67) 1.3 0.98 0.10
M 112 (2.3) 114 (4.22) 1.3 0.98 0.63
H 118 (2.2) 118 (1.46) 1.4 1.00 0.78
300 L 256 (4.0) 199 (17.2) 2.0 1.28 0.01
M 232 (16.0) 272 (1.07) 1.4 0.85 0.01
H 263 (11.7) 258 (1.60) 1.6 1.02 0.50
Metal 30 L 38 (1.8) 40 (1.67) 1.9 0.95 0.28
M 24 (0.9) 31 (0.10) 1.3 0.76 <0.01
H 21 (0.0) 30 (0.06) 1.2 0.71 <0.01
100 L 79 (1.2) 73 (2.32) 1.9 1.08 0.02
M 76 (0.4) 92 (0.65) 1.3 0.83 <0.01
H 79 (0.5) 94 (0.44) 1.2 0.84 <0.01
*

Concentration ranges are: low (L, <103 particles/cm3), medium (M, 103–104 particles/cm3), and high (H, >104 particles/cm3)

**

bold indicates p-value <0.05