Table 3. Comparison of CT values between two real-time PCR platforms using PowerChek kit.

EQA phase Materials Target CT (Mean±SD) P value*
ABI N BioRad N
1st 1,000 upE 28.2 ± 1.3 10 29.0±0.8 13 NS
ORF1a 28.3 ± 1.1 8 28.8±1.1 11 NS
100 upE 31.5 ± 1.2 10 32.0±1.0 13 NS
ORF1a 31.7 ± 0.7 10 32.4 ±1.4 12 NS
10 upE 34.5 ± 1.2 5 36.1±1.2 7 NS
ORF1a 35.4 ± 1.2 5 36.9±2.2 6 NS
2nd A upE 25.5 ± 0.7 14 26.3±0.6 15 0.0011
ORF1a 26.4 ± 0.7 14 26.8±0.7 15 0.0447
B upE 29.2 ± 0.7 14 30.0±0.7 15 0.0035
ORF1a 30.0 ± 0.9 14 30.5±0.8 15 NS

*The comparison of CT values between the two groups was performed by using the Mann-Whitney test because sample sizes of some groups were small.

Abbrviations: EQA, external quality assessment; CT, threshold cycle; NS, not significant.