. 2016 Feb 23;36(3):271–274. doi: 10.3343/alm.2016.36.3.271

Table 1. Accuracy of whole-blood glucose concentration readings using three glucose meter models compared with the measurements obtained by using a chemistry analyzer.

Glucose level (mg/dL) Device Fractional difference between the readings of glucose meters and measurements obtained by using a chemistry analyzer (%)
Vitamin C (mg/dL) Maltose (mg/dL)
0 3 15 30 0 10 40 200 500
Low (60) Roche Accuchek -8.3 10.9 55.0 100.0 -4.9 -0.6 -0.6 1.1 11.4
Barozen H plus -11.9 -5.7 -4.6 -8.2 -6.0 -9.1 -4.7 -9.3 -10.2
Nova Starstrip -14.3 -12.4 -4.6 -10.9 -3.0 -7.1 -7.7 -4.5
Medium (126) Roche Accuchek -8.9 4.3 23.8 44.0 2.9 -8.1 -3.4 1.7 6.5
Barozen H plus -10.5 -10.6 -3.1 -13.0 0.0 -12.8 -4.7 -8.9 -3.4
Nova Starstrip -19.7 -13.9 -13.8 -6.2 -10.6 -10.1 -9.9 -12.2
High (300) Roche Accuchek 4.0 -0.3 11.1 20.4 1.4 -3.2 -5.6 -3.0 0.2
Barozen H plus 7.3 -5.2 -5.2 -7.2 0.8 -5.6 -7.6 -5.7 -9.5
Starstrip -10.9 -12.8 -15.1 -13.4 -11.6 -14.5 -15.7 -16.6 -16.1

The values in bold represent the values that deviated by more than 15% from those determined by a chemistry analyzer.