. 2016 Feb 4;474(4):985–994. doi: 10.1007/s11999-015-4683-1

Table 3.

Correlation between device and patient factors and cumulative rate of volumetric material loss in the CoCr cohort

Variable Spearman’s correlation (ρ) p value
(significant if < 0.05)
Device factors
 Taper angle clearance* [16] 0.06 0.70
 Absolute taper angle clearance* [16] 0.20 0.16
 Head size 0.05 0.72
 Head offset 0.15 0.29
 Lateral offset 0.26 0.07
Stem taper material 0.71 (Kruskal-Wallis)
Stem taper surface finish 0.20 (Mann-Whitney U)
Patient factors
 Implantation time 0.19 0.18
 Patient age at implantation −0.06 0.35
 BMI 0.23 0.07
 UCLA Activity Score 0.15 0.16
 Sex 0.06 (Mann-Whitney U)

* The absolute value of previously estimated taper angle clearance for head-stem junctions, looking at the effect of the net gap on material loss [16].