. 2016 Feb 4;474(4):985–994. doi: 10.1007/s11999-015-4683-1

Table 4.

Reported values of quantified material loss from head-stem tapers in previous studies

Measured
surface
Bearing type Study Number of components Mean volume (± SD)* (range) mm3 Mean rate (± SD)* (range) mm3/year
Female taper CoCr heads
(diameter < 40 mm)
Current Study 50 0.39 (± 0.83) (0–4.34) 0.29 (± 1.24) (0–8.67)
Large-head MOM
(diameter ≥ 40 mm)
Hothi et al. [15] 150 1.52 (0.13–25.89) N/A
Langton et al. [18] 111 N/A Design I: 0.13 (0.01–3.15)
Design II: 0.44 (0.02–8.34)
Bishop et al. [4] 5 8.4 (2.6–20.2) 2.02 (0.6–4.9)
Male taper CoCr heads
(diameter < 40 mm)
Current Study 50 0.10 (± 0.37) (0–2.5) 0.04 (± 0.08) (0–0.32)
Ceramic Current Study 50 0.04 (±0.14) (0–0.74) 0.02 (± 0.08) (0–0.37)
Large-head MOM
(diameter ≥ 40 mm)
Matthies et al. [19] 36 0.29 (0–0.83) 0.08 (0–0.36)
Bishop et al. [4] 2 0.03 (0.02–0.035) 0.01 (0.005–0.006)
Liner backside Large-head MOM
(diameter ≥ 40 mm)
Agne et al. [1] 21 0.4 (0–1.7) 0.2 (0–1.2)

* Added when available; SD available only for the current study; CoCr = cobalt chromium; MoM = metal on metal; N/A = not available.