. 2015 Nov 9;10(2):618–619. doi: 10.1177/1932296815613803

Table 1.

Accuracy Results for the GlucoMen LX Plus System.

Glucose values ≤ 100 mg/dl (5.6 mmol/l)
Glucose values > 100 mg/dl (5.6 mmol/l)
Lot Within ±5 mg/dL Within ±10 mg/dl Within ±15 mg/dl Within ±5% Within ±10% Within ±15%
3211269249 46/64 (72%) 62/64 (97%) 63/64 (98%) 91/146 (62%) 126/146 (86%) 144/146 (99%)
3211325249 40/64 (63%) 55/64 (86%) 60/64 (94%) 92/146 (63%) 139/146 (95%) 144/146 (99%)
3211334249 38/64 (59%) 57/64 (89%) 61/64 (95%) 90/146 (62%) 132/146 (90%) 146/146 (100%)
All 3 lots 124/192 (65%) 174/192 (91%) 184/192 (96%) 273/438 (62%) 397/438 (91%) 434/438 (99%)