Skip to main content
. Author manuscript; available in PMC: 2017 Apr 1.
Published in final edited form as: Ultrasound Med Biol. 2016 Jan 13;42(4):964–970. doi: 10.1016/j.ultrasmedbio.2015.12.004

Table 2.

In situ exposure parameters(mean ± standard deviation) for the Mechanical Index (MI) setting: PCPA, peak compressional pressure amplitude, PRPA, peak rarefactional pressure amplitude, PMPA, peak mean pressure amplitude, and ISPPA, spatial peak, pulse peak intensity.

Setting
MI
PCPA
MPa
PRPA
MPa
PMPA
MPa
Isppa
W cm−2
0.9 3.41 ± 0.29 2.13 ± 0.21 2.77 ± 0.18 170 ± 20
0.7 2.73 ± 0.22 1.71 ± 0.19 2.22 ± .016 114 ± 15
0.52 2.02 ± 0.20 1.34 ± 0.13 1.68 ± 0.14 75 ± 13
0.37 1.42 ± 0.12 1.02 ± 0.11 1.22 ± 0.11 40.5 ± 7.4
0.27 0.98 ± 0.09 0.76 ± 0.08 0.87 ± 0.09 20.8 ± 4.0
0.21 0.70 ± 0.06 0.55 ± 0.06 0.62 ± 0.06 10.4 ± 2.0