Table 3.
Difference in gait measures recorded in ICU survivors between the AMP 331 and manually measured parameters
| Parameter | Walk 1 Mean (95%CI) | Walk 2 Mean (95%CI) |
|---|---|---|
| Distance (m) | -2.79 (-3.66 to -1.92) | -3.11 (-3.99 to -2.22) |
| Step Count (steps) | 0.93 (0.39 to 1.48) | 0.92 (0.44 to 1.40) |
| Speed (cm·s-1) | -21.94 (-29.32 to -14.55) | -28.87 (-34.94 to -22.80) |
There was a consistent underestimation of distance by the AMP331 at all distances during walks one and two (walk 2 mean difference = -3.11m, 95% agreement of – 10.84 to 4.62m).