Table 2. Summary of planar FAIMS and ultra FAIMS devices.
FAIMS Device |
Electrode dimensions |
DV | Carrier gas | Ion Source | Mass Analyzer | Sample | Ref |
---|---|---|---|---|---|---|---|
Sionex SDP-1 microDMX |
L = ~40 mm W = 5 mm Gap = 1.0 mm |
0.3 - 1 kV (3-10 kV/cm) |
N2, Modifiers (e.g., IPA) |
Infused NSI 0.5 μL/min |
Micromass (Waters) ZQ single quadrupole |
Peptide standards | [57,58] |
Homebuilt | L = 50 mm W = 20 mm gap = 2 mm |
3.9 kV (19.5 kV/cm) |
N2, 10-50% He |
Infused ESI 0.4 μL/min |
Agilent TOF with ion funnel |
Peptide standards | [55] |
Homebuilt | L = 50 mm W = 20 mm gap = 1.88 mm |
4.0-5.4 kV (21-28.7 kV/cm) |
N2, 0-70% He |
Infused ESI 0.2 μL/min |
Thermo-Fisher Scientific LCQ ion trap, Thermo-Fisher Scientific LTQ linear ion trap |
Protein standard tryptic digest, peptide standards, phosphopeptides, sequence isomers |
[59-63] |
Owlstone | L = 300 μm W = 115 mm (47 channels) gap = 35 μm |
214 V (61 kV/cm) |
N2 | LC ESI 0.4 mL/min |
Thermo-Fisher Scientific LTQ linear ion trap, Agilent 6230 TOF |
Protein standard tryptic digest, peptide standards |
[65-68] |
DV: dispersion voltage; ESI: electrospray ionization; IPA: isopropyl alcohol; L: length of electrode; LC: liquid chromatography; W: width of electrode