Table 2. Summary of planar FAIMS and ultra FAIMS devices.

FAIMS
Device
Electrode
dimensions
DV Carrier gas Ion Source Mass Analyzer Sample Ref
Sionex
SDP-1
microDMX
L = ~40 mm
W = 5 mm
Gap = 1.0 mm
0.3 - 1 kV
(3-10 kV/cm)
N2,
Modifiers
(e.g., IPA)
Infused
NSI
0.5 μL/min
Micromass
(Waters) ZQ
single quadrupole
Peptide standards [57,58]
Homebuilt L = 50 mm
W = 20 mm
gap = 2 mm
3.9 kV
(19.5 kV/cm)
N2,
10-50% He
Infused
ESI
0.4 μL/min
Agilent TOF with
ion funnel
Peptide standards [55]
Homebuilt L = 50 mm
W = 20 mm
gap = 1.88 mm
4.0-5.4 kV
(21-28.7
kV/cm)
N2,
0-70% He
Infused
ESI
0.2 μL/min
Thermo-Fisher
Scientific LCQ
ion trap,
Thermo-Fisher
Scientific LTQ
linear ion trap
Protein standard
tryptic digest,
peptide standards,
phosphopeptides,
sequence isomers
[59-63]
Owlstone L = 300 μm
W = 115 mm
(47 channels)
gap = 35 μm
214 V
(61 kV/cm)
N2 LC
ESI
0.4 mL/min
Thermo-Fisher
Scientific LTQ
linear ion trap,
Agilent 6230 TOF
Protein standard tryptic digest,
peptide standards
[65-68]

DV: dispersion voltage; ESI: electrospray ionization; IPA: isopropyl alcohol; L: length of electrode; LC: liquid chromatography; W: width of electrode