Table 1.

Sequence parameters used for imaging

Parameter T2-weighted coronal, sagittal DWI coronal T2-weighted DWI
Coil selection Dual: Endovaginal + FLEX-L 2 element phased array body Single: Endovaginal SENSE-XL torso SENSE-XL torso
Sequence Turbo Spin Echo-Multi Shot (TSE factor 10) Spin-Echo Echo Planar Imaging-Single Shot TSE (TSE Factor 13) EPI
TR (ms) 2750 (coronal)
2500 (sagittal)
6500 2100 3230
TE (ms) 80 54 90 53
Flip angle 90° 90° 90° 90°
Bandwidth (Hz/Px) 217.7 9.6 217.3 9.6
FOV (mm) 100 × 100 100 × 100 120 × 200 110 × 100
Acq pixel size (mm) 0.42 × 0.42 1.25 × 1.25 1.0 × 1.0 1.25 × 1.25
Acq matrix 238 80 120 88
Recon matrix 288 224 400 224
Acq voxel size (mm3) 0.35 [0.42 × 0.42 × 2] 3.125 [1.25 × 1.25 × 2] 3.00 [1 × 1 × 3] 4.69 [1.25 × 1.25 × 3]
Slice thickness (mm) 2 (0.1 separation) 2 (0.1 separation) 3 3
Recon voxel size (mm3) 0.24 [0.35 × 0.35 × 2] 0.40 [0.45 × 0.45 × 2] 0.75 [0.5 × 0.5 × 3] 0.72 [0.49 × 0.49 × 3]
Number of slices 24 (coronal)
22 (sagittal)
24 16 16
NSA 2 1 4 2
Included b-values s/mm2
isotropic
n/a 0, 100, 300, 500, 800 n/a 0, 100, 300, 500, 800
Acquisition time 4 min 18 s (coronal)
4 min 0 s (sagittal)
4 min 33 s 3 min 58 s 3 min 50 s
Signal in gel (T1 = 500 ms, T2 = 50 ms) to background noise Flex-L elements 15 cm apart, gel sited centrally within endovaginal coil ring
135
Gel sited centrally within endovaginal coil ring
51
Coil elements 15 cm apart, gel sited centrally
130
Coil elements 15 cm apart, gel sited centrally
7

DWI diffusion-weighted imaging, NSA number of signal averages