Table 1. The results of the evaluation of the proposed algorithm and others based on the AHA ECG database.
File number | TP | FP | FN | SE | PPV |
---|---|---|---|---|---|
Proposed Algorithm | 325679 | 519 | 557 | 99.84 | 99.84 |
Serafim Tabakov, et. al. [15] | 322328 | 2205 | 1085 | 99.32 | 99.66 |
Ivan A Dotsinsky, et. al. [20] | 324739 | 1376 | 1411 | 99.57 | 99.58 |
U. Kunzmann, et. al. [21] | 181564 | 4446 | 309 | 97.61 | 99.83 |
J. Pan, et. al. [14, 23]* | 64404 | 4166 | 8381 | 88.49 | 93.92 |
H. So, et. al. [22, 23]* | 69197 | 623 | 3588 | 95.07 | 99.11 |
*Only 30 records (1201~1210, 2201~2210, 3201~3210) of AHA ECG database were evaluated.