Table 1. The results of the evaluation of the proposed algorithm and others based on the AHA ECG database.

File number TP FP FN SE PPV
Proposed Algorithm 325679 519 557 99.84 99.84
Serafim Tabakov, et. al. [15] 322328 2205 1085 99.32 99.66
Ivan A Dotsinsky, et. al. [20] 324739 1376 1411 99.57 99.58
U. Kunzmann, et. al. [21] 181564 4446 309 97.61 99.83
J. Pan, et. al. [14, 23]* 64404 4166 8381 88.49 93.92
H. So, et. al. [22, 23]* 69197 623 3588 95.07 99.11

*Only 30 records (1201~1210, 2201~2210, 3201~3210) of AHA ECG database were evaluated.