Effect on the MPLM fit of omitting components, using FTIR data from a real sample.

Omitted Cmpnt. ethl. (%) gluc. (%) ac.ac. (%) yst.ex. (%) pep. (%) Y + P (%) ɛRMS (%) 1 − R2 (×10−4) 1 − a (×10−4) b (×10−5)
None 9.52 5.31 0.34 1.04 1.34 2.38 .0569 30 34 7.9
Ethanol 9.23 0.34 0.21 0.00 0.21 .4498 1872 −167 −32
Glucose 15.4 0.76 4.35 0.00 4.35 .4091 1468 480 110
Acetic acid 9.54 5.44 0.98 1.58 2.56 .0792 52 130 31
Yeast extr. 9.57 5.45 0.33 2.34 2.34 .0633 36 60 14
Peptone 9.36 5.18 0.36 2.31 2.31 .0671 41 26 6.1
Y + P 7.28 6.54 0.77 .2299 380 480 120