Table 2.
Precision or coefficient of variation (CV) reflecting phe and tyr measurement with three different methods.
| Phe |
Tyr |
||||||
|---|---|---|---|---|---|---|---|
|
Control level |
n |
Concentration (μM) mean ± SD |
CV (%) |
n |
Concentration (μM) mean ± SD |
CV (%) |
|
| Plasma-IEC | Low | 20 | 79 ± 1 | 1.6% | 20 | 61 ± 3 | 4.9% |
| High | 20 | 403 ± 7 | 1.6% | 20 | 223 ± 13 | 5.9% | |
| DBS-IEC | Low | 20 | 32 ± 2 | 5.9% | 20 | 38 ± 4 | 9.2% |
| High | 20 | 482 ± 17 | 3.5% | 20 | 462 ± 23 | 5.1% | |
| DBS-MS/MS | Low | 20 | 33 ± 4 | 11.2% | 20 | 33 ± 3 | 10.3% |
| High | 20 | 371 ± 43 | 11.7% | 20 | 648 ± 79 | 12.1% | |