Table 5. Comparison of the strength of structure-function relationship between visual field measurement instruments.
location | Statistic | MAIA | MP-3 | HFA |
---|---|---|---|---|
Global | Correlation coefficient | 0.627 | 0.633 | 0.682 |
95% bootstrapped CI | 0.515–0.750 | 0.526–0.732 | 0.601–0.767 | |
Superior henifield | Correlation coefficient | 0.547 | 0.567 | 0.602 |
95% bootstrapped CI | 0.419–0.676 | 0.454–0.672 | 0.508–0.721 | |
Inferior hemifield | Correlation coefficient | 0.656 | 0.687 | 0.642 |
95% bootstrapped CI | 0.551–0.756 | 0.583–0.774 | 0.585–0.757 | |
Superotemporal | Correlation coefficient | 0.592 | 0.603 | 0.672 |
95% bootstrapped CI | 0.447–0.691 | 0.492–0.714 | 0.570–0.772 | |
Superior | Correlation coefficient | 0.426 | 0.46 | 0.517 |
95% bootstrapped CI | 0.275–0.560 | 0.333–0.577 | 0.382–0.632 | |
Superonasal | Correlation coefficient | 0.417 | 0.424 | 0.408 |
95% bootstrapped CI | 0.265–0.572 | 0.272–0.554 | 0.257–0.548 | |
Inferotemporal | Correlation coefficient | 0.673 | 0.691 | 0.718 |
95% bootstrapped CI | 0.523–0.767 | 0.590–0.784 | 0.623–0.807 | |
Inferior | Correlation coefficient | 0.631 | 0.67 | 0.651 |
95% bootstrapped CI | 0.539–0.719 | 0.569–0.755 | 0.560–0.730 | |
Inferonasal | Correlation coefficient | 0.579 | 0.616 | 0.587 |
95% bootstrapped CI | 0.452–0.716 | 0.509–0.718 | 0.485–0.681 |
CI: confidential interval.