Skip to main content
. 2015 May 22;75(4):1717–1729. doi: 10.1002/mrm.25767

Table 1.

Overview of Typical Sequence Parameters Used in This Study

Parameter Clinical T2* (cardiac) Clinical T2* (liver) Phantom measurements
T2* sGRE (reference) T2* mGRE (cardiac) T2* mGRE (liver) T1 MOLLI
Acquired voxel size [mm] 2 × 2 3 × 3 1.96 × 2.0 1.96 × 2.15 1.96 × 2.0 1.98 × 2.0
Slice thickness [mm] 10 10 8 10 10 10
Matrix size 160 116 112 × 106 112 × 102 112 × 110 116 × 90
FOV [mm] 320 × 320 348 × 348 220 × 212 220 × 220 220 × 220 230 × 180
TEs [ms] 2.5, 5.0, 7.5, 10.0, 12.5, 15.0, 17.5, 20.0, 22.5 and 25.0 1.2, 2.7, 4.2, 5.7, 7.2, 8.7, 10.2, 11.7, 13.2 and 14.7 1.34, 2, 3, 5, 7.5, 10, 12.5, 15, 20, 30, 40, 50, 75, 100, 150, 200, 300 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20, 22.5 and 25.0 1.3, 3.4, 5.5, 7.6, 9.7, 11.8, 13.9, 16, 18, 20.1 1.11
TR [ms] 26 ms 17 ms 6 × T1 26 38 2.4
Flip Angle 20° 20° 50° 20° 20° 35°
Parallel imaging Factor 2 (SENSE) No No Factor 2 (SENSE) No Factor 2 (SENSE)
Read‐out profile Linear Linear Linear Linear Linear Linear
Turbo factor 6 No turbo factor; No turbo factor; 6 No turbo factor; 48
Preparation pulse: DIR SPIR No preparation DIR SPIR IR
Flow‐compensation On Off Off On Off On
T1 mapping scheme: n/a n/a n/a n/a n/a 5(3s)3;