Table 1. Computed tomographic scan parameters.
Parameters | |
---|---|
Scanner | Siemens SOMATOM Sensation 64 |
Scan type | Helical multi detector row CT angiography |
Slice thickness | 64 detector row ×0.6 mm collimator width |
Helical detector pitch | 0.9 |
Gantry rotation speed | 0.37 s |
Tube potential | 120 kV |
Tube current | 180 mA |
Contrast | Omnipaque 350 100 mL IV injection 4 mL per second |
Scanning range | Pubic symphysis to 4 cm above umbilicus |
Scanning direction | Caudo-cranial |
Bolus tracking | +100 HU from common femoral artery with minimal delay |
Automatic dose modulation (Siemens Care Dose 4D) | Disabled |
Imaging reconstruction | 1 mm/0.7 mm overlapping axial images |
CT, computed tomographic; HU, Hounsfield units.