Tab. 1.

Representative examples of coefficients of variation for sIgE measurements against single allergens depending on signal strength

Sample Allergen Signal strength ISU-E Mean ISU-E CV intra-assay variance (%) CV inter-assay variance (%)
1 rPar j 2 0.33–0.98 0.32 18 9
2 nGal d 1 0.46 11 16
3 nCry j 1 0.98 12 13
4 rEqu c 1 1.2–14 1.2 15 11
5 rDer f 1 4.6 5 9
6 rFel d 1 14 8 9
7 rAra h 1 19–90 19 11 13
8 rPhl p 5b 47 6 7
9 rBet v 1 90 7 7

CV, coefficient of variation