WNV NAT RealStar WNV RT-PCR Kit in combination with chemagic viral 2K (from 2.4 ml plasma) extraction protocol

WNV-RNA concentration, copies/mla Reactive samples/number tested Percent positive
179 24/24 100
89.5b 24/24 100
44.8 24/24 100
22.4 19/24 75
11.2 12/24 50
5.6 5/24 21

WNV-RNA concentration, copies/ml

95 % detection limit (95% CI) 30.2 (24.2-45.4)
Intra-assay WNVb, mean ±SD / CV 33.20 ± 0.42 / 1.27
Interassay WNVb, mean ±SD / CV 32.62 ± 0.86 / 2.63
a

NATtrol(tm) WNV (NATWNV-0005, lot number 309897); calibrated to Health Canada WNV Reference preparation.

b

Intra- and interassay variabilities were calculated exemplarily for the concentration of 89.5 copies/ml.