Skip to main content
. 2016 Mar 17;43(4):1720–1735. doi: 10.1118/1.4943632

TABLE III.

Linear fit parameters for calculating image-based attenuation coefficients.

Grid-in Grid-out
Tube voltage (kV) m (voxel value/cm−1) k (voxel value) m (voxel value/cm−1) k (voxel value)
28 6688 613 6368 881
29 6720 617 6416 856
31 6788 630 6483 828