Skip to main content
. 2016 Mar 22;11(3):e0151247. doi: 10.1371/journal.pone.0151247

Table 4. Procedural data.

All patients 23mm 26mm 29mm P-Value
Number of patients, N 235 77 91 67
Nongeneral anesthesia 235 (100%) 77 (100%) 91 (100%) 67 (100%)
General anesthesia 0 (0%) 0 (0%) 0 (0%) 0 (0%)
Correct placement at intended site 235 (100%) 77 (100%) 91 (100%) 67 (100%)
Cardiopulmonary bypass 0 (0%) 0 (0%) 0 (0%) 0 (0%)
Aortic regurgitation after valve placement by aortography
 None / trace 219 (93.2%) 73 (94.8%) 81 (89.0%) 65 (97.0%) 0.12
 Mild 16 (6.8%) 4 (5.2%) 10 (10.9%) 2 (3.0%)
 Moderate 0 0 0 0
 Severe 0 0 0 0
Balloon post-dilation 0 0 0 0
Implantation of >1 valve 0 0 0 0
Adjunctive PCI 0 0 0 0
Coronary obstruction 0 0 0 0
Annular rupture 0 0 0 0
Conversion to surgery 0 0 0 0
Aortic regurgitation by echocardiography post TAVR
 None / trace 208 (88.5%) 67 (87.0%) 80 (87.9%) 61 (91.0%) 0.73
 Mild 27 (11.5%) 10 (12.9%) 11 (12.1%) 6 (8.9%)
 Moderate 0 0 0 0
 Severe 0 0 0 0
Mean regurgitation index 24.2 ± 8.9 22.7 ± 8.5 24.7 ± 8.3 25.3 ± 10.0 0.25
Mean aortic gradient, mmHg 12.6 ± 5.0 15.4 ± 5.6 12.0 ± 4.2 10.2 ± 3.6 <0.001
Device success 228 (97.0%) 72 (93.5%) 89 (97.8%) 67 (100%) 0.05
Contras amount, mL 86 ± 24 83 ± 28 91 ± 21 85 ± 23 0.11

Values are mean ± SD or n (%); PCI = percutaneous coronary intervention