Important aortic measurement, which identify the optimal CoreValve trans catheter heart valve [38] (http://www.accessdata.fda.gov/cdrh_docs/pdf13/P130021c.pdf)

Valve Aortic Annulus Diameter (mm) Aortic Annulus Area (mm2) Aortic Annulus Perimeter Ascending Aortic Diameter Sinus of valsalva: Height/Width (mm)
CoreValve 23 18–20 254.5–314.2 56.5–62.8 ≤34 mm ≥15/≥25
CoreValve 26 20–23 314.2–415.5 62.8–72.3 ≤40 mm ≥15/≥27
CoreValve 29 23–26 415.5–572.6 72.3–84.8 ≤43 mm ≥15/≥29
CoreValve 31 26–29 530.9–660.5 81.7–91.1 ≤43 mm ≥15/≥29