Table 2.

Suggested MRI pulse sequence protocol at 3T and (1.5T)

3T (1.5T) Coronal STIR wide FOV Axial PD wide FOV Sagittal PD Axial oblique PD Coronal PD
TR/TE (ms) 4000/19 (4000/17) 4800/40 (4000/30) 4000/30 (3500/26) 4000/40 (4000/26) 4000/32 (4000/26)
Flip angle 180 (180) 180 (180) 180 (180) 180 (180) 180 (180)
ETL 12 (7) 12 (7) 12 (8) 12 (9) 12 (7)
RBW (kHz) 50 (32) 50 (32) 50 (32) 50 (32) 50 (32)
NEX 2 (2) 1 (2) 2 (3) 2 (3) 2 (3)
Matrix 288 × 288 (256 × 192) 512 × 256 (512 × 256) 512 × 384 (512 × 384) 512 × 320 (512 × 256) 512 × 384 (512 × 384)
FOV (cm) 38 (34) 36 (36) 19 (20) 19 (20) 19 (20)
Slice thickness (mm)/gap 5.5/0 (5.0/0) 5.0/0 (5.0/0) 2.5/0 (2.5–2.8/0) 3.0/0 (3.0/0) 3.0 (3.0/0)

STIR short tau inversion recovery, PD proton density, TR repetition time, TE echo time, ETL echo train length, RBW receiver bandwidth, NEX number of excitations, FOV field of view