Table 4.
References | Number of hip implants | Implant description | Results/findings |
---|---|---|---|
Banerjee et al. [2] | 5 | Five different stem designs | Trunnions exhibited either a fracture or gross loss of material volume upon revision |
Hohman et al. [34] | 1 | First-generation proximally coated titanium cementless stem | Damaged liner insert, an intact femoral head, diffuse metallic debris, and excessive wear of the trunnion |
Pansard et al. [35] | 1 | 12-mm/14-mm trunnion | Abnormal mobility between the femoral head and the trunnion, with severe damage at the taper. Macroscopic pitting and wear |