Table 3.

Comparative performance of positron emission tomography-magnetic resonance imaging maximum standardized uptake value and minimum apparent diffusion coefficient for lesion detection compared with positron emission tomography-computed tomography reference

Test characteristic SUVmax > 2.5 ADCmin < 0.5 mm2/s ADCmin < 1.0 mm2/s ADCmin < 1.5 mm2/s
Accuracy 100%1 61.0% 68.3% 58.5%
Sensitivity 100% 28.6% 70.0% 100.0%
Specificity 100% 95.0% 66.7% 19.0%
Positive predictive value 100% 85.7% 66.7% 54.1%
1

Indicates statistical significance (P < 0.001, two-tailed McNemar’s test). ADCmin: Minimum apparent diffusion coefficient; SUVmax: Maximum standardized uptake value.