TABLE 1.

Summary of sensitivities, specificities, positive predictive values, and negative predictive values of HIV rapid tests in comparison to those of EIA plus Multispot and EIA alone in Panama

Test Totala Result (n)b
Sensitivity, (% [95% CI]) Specificity (% [95% CI]) Predictive value (%)c
TP FP TN FN Positive Negative
Determine 1,503 465 0 1,037 1 99.8 (98.81–99.99) 100.0 (99.64–100) 100.0 99.9
Uni-Gold 1,508 467 0 1,039 1 99.8 (98.47–99.95) 100.0 (99.65–100) 100.0 99.8
SD Bioline 1,509 468 0 1,040 1 99.8 (98.82–99.99) 100.0 (99.65–100) 100.0 99.9
INSTI 1,505 467 2 1,035 1 99.8 (98.82–99.99) 99.8 (98.30–99.98) 99.6 99.9
a

Total numbers vary due to an inadequate volume of specimen or invalid results on a given rapid test.

b

TP, true positive; FP, false positive; TN, true negative; FN, false negative.

c

Predictive values are a function of HIV prevalence, so 95% CI are not included.