Table 1.
Comparison of the corneal power and WTW distance values measured by the 2 devices
Parameters | Orbscan II | iTrace | P valuea |
---|---|---|---|
Kf (D) | 42.56 ± 1.34 | 42.06 ± 1.34 | <0.0001 |
Ks (D) | 43.76 ± 1.58 | 43.21 ± 1.57 | <0.0001 |
WTW (mm) | 11.57 ± 0.34 | 11.33 ± 0.36 | <0.0001 |
Data are expressed as mean ± SD; apaired t test <0.05 considered significant; Kf = flat axis power; Ks = steep axis power, D = diopter