Table 2.

Method validation data using different internal standard options. Accuracy is expressed as the percent measurement in comparison to true concentration. Precision is shown as the percent CV for replicate analyses over three days of measurement. The SIL-VVV extended refers to the extended peptide, digest refers to adding the internal standard prior to digestion and quench to adding the internal standard after trypsin digestion. For RDH11, the mid-QC was prepared as a pooled sample of human liver microsomes from four donors.

Internal Standard QC (low) QC (mid) QC (high) r2 for calibra-
tion curve
Accuracy (%) CV% Accuracy (%) CV% Accuracy (%) CV%
RDH11 22.5 nM 2 mg/mL pooled HLM (n=4) 55 nM
SIL-RDH 89 17 16 81 15 0.97± 0.01
SIL-VVV extended 153 16 12 122 11 0.91± 0.10
SIL-VVV in digestion 120 10 13 113 15 0.96± 0.01
SIL-VVV quench 98 8 8 94 8 0.98± 0.01
ALDH1A1 150 nM 325 nM 600 nM
SIL-ANN-1A1 in digestion 74 33 93 23 127 20 0.96± 0.02
SIL-ANN-1A1 quench 110 10 111 14 111 11 0.99± 0.01
ALDH1A2 10 nM 32.5 nM 80 nM
ILE in digestion 109 26 120 4 116 16 0.96± 0.02
ILE quench 105 14 92 11 89 15 0.98± 0.01