Table 1.

Fold-change measurement error.

Target gene expression level (Cq)
<30 30–32.5 >32.5
No PreAmp Average fold change measurement error (Cq) 0.093 0.265 0.804
Average fold change measurement error (percent) 7% 20% 75%
88-plex PreAmp Average fold change measurement error (Cq) 0.094 0.162 0.578
Average fold change measurement error (percent) 7% 12% 49%
Statistical equivalence with no PreAmp. p-value (upper threshold) <0.0001a 0.0121a 0.0515
Statistical equivalence with no PreAmp. p-value (lower threshold) 0.0021a 0.0003a <0.0001a
355-plex PreAmp Average fold change measurement error (Cq) 0.108 0.119 0.380
Average fold change measurement error (percent) 8% 9% 30%
Statistical equivalence with no PreAmp. p-value (upper threshold) <0.0001a 0.0010a 0.0074a
Statistical equivalence with no PreAmp. p-value (lower threshold) 0.0205a 0.0050a 0.0002a
Estimated fold-change limits-of-detection 2-fold 4-fold 8-fold
a

Statistically equivalent with no PreAmp.