Table 2.

Confirmation ion ratios for a candidate reference measurement procedure for the quantitation of 25(OH)D2 and 25(OH)D3 in calibrators and serum samples

n Concentration,
nmol/L
Mean Ion Ratioa
(range)

25(OH)D2
Calibrator 40 3.71 0.57 (0.45–0.68)b
Serum samples 67 LOQ-64.9 0.56 (0.45–0.69)c
25(OH)D3
Calibrators 80 10–100 0.57 (0.49–0.66)d
Serum samples 148 16–183 0.57 (0.51–0.63)c
a

Peak area ratios of confirmation (m/z 383.3→105.0) to quantitation (m/z 383.3→365.1) ions

b

Mean (± 20%) calculated from calibration level 2 in 10 measurement series (2 preparations, 2 injections)

c

Mean (minimum-maximum) from all serum samples with concentrations >LOQ from 5 measurement series

d

Mean (± 15%) calculated from all calibration levels in 5 measurement series (4 levels, 2 preparations, 2 injections)