Table 3.

Imprecision of the candidate reference measurement procedure for quantitation of serum concentrations of 25(OH)D2 and 25(OH)D3 in five internal quality control materialsa

Concentration, nmol/L CV, %
Within-day Between-day Total
25(OH)D2
3.00 1.6 2.8 3.1
3.08 3.8 3.7 4.9
4.66 3.4 3.0 4.1
9.90 1.5 2.9 3.1
14.8 1.7 2.1 2.5
25(OH)D3
23.8 2.3 1.2 2.2
40.6 3.1 1.0 2.7
49.2 0.6 1.5 1.6
65.3 1.8 0.8 1.7
84.5 1.5 2.0 1.8
a

Duplicate samples of internal quality control serum were tested in five independent measurement series. The within-day, between-day imprecision, and total CV were calculated for each material using CLSI EP-10-A3 document as a guide [24].