Table 1.

Dose Rate on Central Axis of the Field at Varying Depths in the Solid-Water Phantom Measured by EBT3 Film for 10 Collimators.

Dose Rate on Central Axis, cGy/min
Square Collimators, mm2
Circular Collimators, Diameter in mm
Depth, mm 10 × 10 20 × 20 40 × 40 ϕ1 ϕ2.5 ϕ3.5 ϕ5 ϕ10 ϕ15 ϕ25
1 324.5 339.2 364.6 177.8 239.8 236.9 310.2 320.9 329.8 344.5
5 309.7 323.7 360.2 169.9 225.7 222.2 285.4 305.3 321.2 328.5
10 291.2 310.3 341.1 162.0 208.7 207.3 269.1 285.4 291.7 315.4
20 240.0 273.2 309.8 128.6 173.1 171.6 216.6 236.2 251.2 271.1
30 205.5 226.1 273.7 103.7 144.6 147.4 180.0 200.1 209.2 239.5
40 170.5 192.2 235.5 90.5 118.2 121.1 154.1 163.8 179.5 197.9
50 141.5 174.0 198.8 73.9 103.4 98.3 126.1 139.6 153.3 170.7