Table 2.

The Dose Difference Between Metropolis Calculation and Film Measurement.

Collimator, mm2 or mm in Diameter Dose Rate Measured With EBT3 Film, cGy/min Dose Rate Calculated in Metropolis, cGy/min % Difference in Dose
10 × 10 240.0 233.0 –2.9
20 × 20 273.2 267.5 –2.1
40 × 40 309.8 305.6 –1.4
ϕ1 128.6 126.2 –1.8
ϕ2.5 173.1 169.2 –2.3
ϕ3.5 171.6 166.7 –2.8
ϕ5 216.6 214.6 –0.9
ϕ10 236.2 232.0 –1.8
ϕ15 251.2 247.9 –1.3
ϕ25 271.1 268.0 –1.1