Table 1.
BioDent | OsteoProbe | |
---|---|---|
Method of use | Mounted* | Hand-held |
Manufacturer suggested research target | Pre-clinical | Clinical |
Reference probe | Three types | None |
Test probe size | ~370 microns | ~ 370 microns |
Probe tip radius | 2.5 microns | <10 microns |
Overcoming soft tissue | Scraping, preload, preconditioning | Unknown |
Loading force** | User defined (2–10 N) | 40 N |
Loading rate | 40 N/sec | 120,000 N/sec |
Loading cycles | User defined (up to 20) | 1 |
Preconditioning | Optional in cycle number and force | During 10N preload |
Data visualization during test | Force versus distance plots and data | Raw (unnormalized) BMSi value |
Although the majority of studies have used the system mounted, data does exist using the BioDent as a hand-held device(17).
Biodent uses a true load-control feedback system while Osteoprobe generates a load using a spring-based system without feedback.