Table 1.

Characteristics of the two reference point indentation devices

BioDent OsteoProbe
Method of use Mounted* Hand-held
Manufacturer suggested research target Pre-clinical Clinical
Reference probe Three types None
Test probe size ~370 microns ~ 370 microns
Probe tip radius 2.5 microns <10 microns
Overcoming soft tissue Scraping, preload, preconditioning Unknown
Loading force** User defined (2–10 N) 40 N
Loading rate 40 N/sec 120,000 N/sec
Loading cycles User defined (up to 20) 1
Preconditioning Optional in cycle number and force During 10N preload
Data visualization during test Force versus distance plots and data Raw (unnormalized) BMSi value
*

Although the majority of studies have used the system mounted, data does exist using the BioDent as a hand-held device(17).

**

Biodent uses a true load-control feedback system while Osteoprobe generates a load using a spring-based system without feedback.