In vivo quantitative measurement by OFDI and IVUS at baseline, 4, and 12 weeks

Baseline 4 weeks 12 weeks
Total number of stents: 34 17 13
Coronary artery (LAD/LCx/RCA): 12/11/11 7/3/7 5/4/4
OFDI IVUS P value OFDI IVUS P value OFDI IVUS P value
Mean LD (mm) 2.65 ± 0.26 2.51 ± 0.24 <0.001 2.31 ± 0.42 2.30 ± 0.35 0.96 1.86 ± 0.34 1.94 ± 0.32 0.002
Minimum lumen area (mm2) 5.15 ± 0.96 4.73 ± 0.95 <0.001 3.29 ± 1.30 3.86 ± 1.15 0.001 2.25 ± 0.87 2.48 ± 0.82 0.001
Lumen VI (mm3/mm) 5.50 ± 1.11 4.95 ± 0.99 <0.001 4.27 ± 1.46 4.22 ± 1.26 0.49 2.76 ± 0.93 3.00 ± 0.96 0.002
Minimum stent area, tracing at LE (mm2) 4.63 ± 0.89 4.73 ± 0.95 0.02 5.05 ± 1.21 4.81 ± 1.09 0.002 4.69 ± 0.83 4.83 ± 0.86 0.03
Minimum stent area, tracing at HIP (mm2) 4.92 ± 0.94 4.73 ± 0.95 <0.001  5.34 ± 1.27 4.81 ± 1.09 <0.001 4.98 ± 0.87 4.83 ± 0.86 0.03
Stent VI, tracing at LE (mm3/mm) 4.82 ± 0.92 4.95 ± 0.99 <0.001 5.20 ± 1.21 4.95 ± 1.15 <0.001 4.87 ± 0.89 4.98 ± 0.92 0.08
Stent VI, tracing at HIP (mm3/mm) 5.07 ± 0.96 4.95 ± 0.99 0.001 5.49 ± 1.27 4.95 ± 1.15 <0.001 5.16 ± 0.93 4.98 ± 0.92 0.01
Neointimal VI, tracing at LE (mm3/mm) N/A N/A N/A 0.93 ± 0.45 0.73 ± 0.32 0.002 2.11 ± 0.55 1.98 ± 0.63 0.01
Neointimal VI, tracing at HIP (mm3/mm) N/A N/A N/A 1.22 ± 0.43 0.73 ± 0.32 <0.001 2.40 ± 0.58 1.98 ± 0.63 <0.001

HIP highest intensity points, IVUS intravascular ultrasound, LAD left anterior descending artery, LCx left circumflex artery, LD lumen diameter, LE leading edge, OFDI optical frequency domain imaging, RCA right coronary artery, VI volume index