Skip to main content
. 2015 Aug 14;31:79–88. doi: 10.1007/s12928-015-0349-x

Table 4.

Intra- and interobserver variability in assessing swine coronary segments by OFDI and IVUS

Intraobserver variability Interobserver variability
Mean difference between 1st and 2nd time analysis Coefficient of variation (%) Mean difference between 1st and 2nd observer Coefficient of variation (%)
OFDI
 Mean LD (mm) 0.02 ± 0.02 1.1 0.04 ± 0.02 0.9
 Minimum lumen area (mm2) 0.06 ± 0.08 2.6 0.14 ± 0.06 2.0
 Lumen VI (mm3/mm) 0.07 ± 0.09 2.4 0.12 ± 0.09 2.3
 Minimum stent area, tracing at LE (mm2) −0.03 ± 0.06 1.4 −0.04 ± 0.07 1.6
 Minimum stent area, tracing at HIP (mm2) −0.05 ± 0.09 1.9 0.01 ± 0.07 1.4
 Stent VI, tracing at LE (mm3/mm) −0.03 ± 0.04 0.9 −0.09 ± 0.04 0.8
 Stent VI, tracing at HIP (mm3/mm) −0.04 ± 0.04 0.8 −0.03 ± 0.05 1.0
 Neointimal VI, tracing at LE (mm3/mm) −0.05 ± 0.06 3.0 −0.16 ± 0.05 2.4
 Neointimal VI, tracing at HIP (mm3/mm) −0.07 ± 0.07 3.0 −0.09 ± 0.07 3.5
IVUS
 Mean LD (mm) −0.02 ± 0.09 4.1 0.003 ± 0.11 5.4
 Minimum lumen area (mm2) −0.10 ± 0.47 14.1 0.02 ± 0.53 16.3
 Lumen VI (mm3/mm) −0.08 ± 0.26 7.2 −0.002 ± 0.34 9.5
 Minimum stent area (mm2) 0.01 ± 0.10 2.1 0.18 ± 0.16 3.4
 Stent VI (mm3/mm) 0.00 ± 0.06 1.2 0.001 ± 0.03 0.6
 Neointimal VI (mm3/mm) 0.12 ± 0.34 19.4 0.005 ± 0.42 22.9

Second observation by same investigator was performed at least 4 weeks later after 1st observation

HIP highest intensity points, IVUS intravascular ultrasound, LD lumen diameter, LE leading edge, OFDI optical frequency domain imaging, VI volume index