Skip to main content
. 2016 Apr 12;6:80. doi: 10.3389/fonc.2016.00080

Table 1.

Comparison of the properties of the different block detectors from Siemens Medical Solutions USA, Inc. Molecular Imaging Division, whose sketch is depicted in Figure 6.

Detector name LSO2 BGO1
Active volume (mm3) 52.7 × 52.7 × 20.0 52.7 × 52.7 × 20.0
Granularity Independent pixels Segmented crystal
Pixel matrix 13 × 13 8 × 8
Pixel dimensions (mm3) 4.0 × 4.0 × 20.0 6.4 × 6.4 × 20.0
Light guide Coupled to block Cut into block
Operating voltage +900 V +1350 V
Commercial scanner Biograph PET/CT ECAT EXACT 47 PET

Spatial dimensions are given as height × width × depth. Reproduced with permission from Ref. (53).