Skip to main content
. 2016 Mar 13;8(3):e531. doi: 10.7759/cureus.531

Table 2. Summary of systematic and random positional errors stratified by region of interest used for CBCT image registration.

  Target Nose
  LR (mm) CC (mm) AP (mm) LR (mm) CC (mm) AP (mm)
Localization CBCT
Group Mean -0.54 0.06 -1.24 -0.83 -0.33 -1.80
Systematic Error 0.47 1.18 0.50 0.52 1.74 0.28
Random Error 0.78 1.08 0.53 0.88 1.29 0.52
Verification CBCT
Group Mean 0.04 -0.18 -0.02 -0.31 -0.53 -0.63
Systematic Error 0.16 0.14 0.17 0.47 0.81 0.68
Random Error 0.32 0.34 0.32 0.46 0.65 1.09
Post-Treatment CBCT
Group Mean 0.04 -0.06 -0.19 -0.33 -0.36 -0.68
Systematic Error 0.25 0.18 0.22 0.53 0.70 0.48
Random Error 0.38 0.43 0.40 0.48 0.58 0.47