Skip to main content
. 2016 Jul 15;81:166–172. doi: 10.1016/j.bios.2016.02.059

Fig. 4.

Fig. 4.

Analysis of in vitro bladder model conditions. (a) pH of residual bladder model urine at model start (0 h), on initial colour change (average 4.2 h), maximal colour change (average 6.2 h), and on catheter blockage (average 16.1 h) (b) Measured fluorescence of bladder model urine at the same time points. **** p<0.0001. Data shown is the mean of triplicate repeats. Error bars represent standard error of the mean (SEM). For part B only ## p<0.01; #### p<0.0001P. mirabilis at 0 h vs P. mirabilis at 6.2 h.