Published in final edited form as: Int J Radiat Oncol Biol Phys. 2016 Feb 10;95(1):242–248. doi: 10.1016/j.ijrobp.2016.01.061

Table 2.

Differences between the Treatment Planning System dose calculations and the TLD and film gamma analyses.

HEAD LIVER LUNG PROSTATE SPINE
TLD
ECLIPSE 0.99 0.95 0.97 0.96 0.98
IN-HOUSE 1.01 0.93 0.95 0.99 0.94
RAYSTATION 0.97 0.97 0.95 0.98 -
XIO 1.00 0.96 0.96 0.96 0.98
ANOVA P-VALUE 0.072 0.214 0.133 0.006* 0.035*
Gamma
ECLIPSE 97 78 91 93 N/A
IN-HOUSE 96 86 71 87 N/A
RAYSTATION 96 98 84 99 N/A
XIO 97 91 87 89 N/A
KRUSKAL-WALLIS P-VALUE 0.692 0.096 0.048* 0.015* N/A
*

Indicates analysis was statistically significant with p<0.05.