TABLE I.

Cryoablation Algorithms for Two Treatment Systems Used

Parameter Cryoablation System and Maximal Tumor Diameter
Visicaa Visica 2a

<10 mm 10–15 mm <10 mm 10–15 mm
Probe tip distance beyond lesion (mm) 12–14 9–12 20–24 17–20
Duration of first freeze (min)b 8 10 6 8
Duration of thaw (min) 10 10 10 10
Duration of second freeze (min)b 8 10 6 8
a

Sanarus Medical.

b

Freeze time added to achieve ice ball diameter goal.