Skip to main content
. 2016 Apr 19;43(5):2399–2409. doi: 10.1118/1.4946816

TABLE II.

MTF-10 along the tangential direction for all scan protocols and off-centered positions included in this study.

Tangential MTF-10 (mm−1)
Conventional Hi-Res
Kernel 0 cm 4 cm 8 cm 12 cm 16 cm 0 cm 4 cm 8 cm 12 cm 16 cm
Soft 0.62 0.62 0.60 0.55 0.48 0.62 0.63 0.62 0.61 0.58
Standard 0.71 0.70 0.66 0.59 0.50 0.71 0.71 0.70 0.68 0.63
Detail 0.84 0.83 0.75 0.63 0.52 0.86 0.86 0.82 0.75 0.67
Lung 0.90 0.89 0.85 0.78 0.59 0.89 0.91 0.88 0.85 0.77
Chest 1.08 1.06 0.91 0.75 0.56 1.42 1.39 1.23 0.95 0.78
Bone 1.21 1.17 0.99 0.78 0.57 1.23 1.23 1.14 0.94 0.78
Bone plus 1.21 1.18 1.02 0.84 0.60 1.22 1.23 1.15 0.97 0.82
Edge 1.46 1.42 1.12 0.83 0.58 1.46 1.44 1.29 0.97 0.80
HD Standard 0.84 0.83 0.80 0.77 0.69
HD Detail 1.05 1.05 0.97 0.85 0.71
HD Lung 1.12 1.13 1.06 0.94 0.81
HD Bone 1.38 1.37 1.24 0.97 0.79
HD Bone plus 1.59 1.54 1.28 0.97 0.80
HD Ultra 1.74 1.70 1.38 1.04 1.09
HD Edge 1.95 1.93 1.41 1.09 0.91